NewSonic SonoDur3超聲波硬度計(jì)
The New UCI
Generation after SonoDur2
Mobile Hardness testing onmetals with UCI *
in productionand maintenance. Starting withgoods income inspection untilthe fi
nal product is reached.
*Ultrasonic Contact Impedance
Norms: ASTM A 1038 and DIN 50159-1, 2
Conversions: ASTM E 140-13, EN ISO 18265
Multiple Solutions for theSonoDur Family
and also forMIC10/MIC20 (by probe retrofi t)
NewSonic SonoDur3超聲波硬度計(jì)
Motor
probes (HV 0,1 – HV 0,8): 0.1 kgf, 0.3 kgf, 0.8 kgf
Handheld
probes (HV 1 – HV 10): 1 kgf, 3kgf, 5 kgf, 10 kgf
The better Way to
Hardness Testing
Rugged,
easy to use, fast and precise. Made in Germany.
Instrument
Features
? Carrying strap, edge protection, IP65,
MIL-810G (vibration, shock, drop)
? Brilliant color display in 5’’, smooth
glass shield for easy cleaning
? Direct function access just by touch,
intuitive menu control, ideal forMIC10/MIC20 users and others
? No set-up needed at beginning of shift –
ready for immediate use
NewSonic SonoDur3超聲波硬度計(jì)
Android
Operating System and Data Transfer
? USB: transfer of measurement data in
txt-format using thefile explorer
? WLAN: automatic transfer of single values, completed data stets
? Bluetooth: manual transfer of a set of
measurement data
?“Unlimited”storage capacity for
measurement data and settings
Intelligent
Management ofMeasurement Data and Settings
? Simple access to data and qualified
back-traceableassessment of results
? File names can be predefined, automatic
closingof data sets
?“AllMeas”- summarized list of results in
storeddata sets
Reliable
MeasurementTechnique with long-Term Stability
Efficient working tool, optimized for the
daily needs of the inspector
Unique
Can be calibrated in full according to
DAKKS after DIN 50159-2 (calibration
curve 150 – through 900 HV, diamond,
nominal test force)
NewSonic SonoDur3超聲波硬度計(jì)
Measuring
Specifications
|
Measuring
principle
|
UCI Method, corresponds to DIN 50159,
ASTM A1038
|
|
Test
indenter
|
Vickers diamond 136°
|
|
Test
loads
Newton scale (1kgf = 9.81 N)
|
Motor probes: 1N (0.1 kgf), 3N (0.3kgf)
and 8.6 N (0.9 kgf)
Handheld Probes: 10N (1 kgf), 30N (3kgf),
49N (5kgf), 98N(10kgf), (Other test loads on request)
|
|
Hardness scales and
range
(according to relevant standards), in this
casetable A1 respectively T1, T2 (low alloy steel).
Different measuring ranges are valid for
othermaterials. When exceeding the limits theconversion range will be
extended. The calcu-lated values are highlighted in red besides theoriginal
data in HV.
Note:
Conversions are acc. to latest ASTM
E140-12b?1(2013) und EN ISO
18265:2014.
Conversions into tensile strength: 98N
(10kgf)test load only.
|
Vickers
Brinell
Rockwell
Rockwell
Rockwell
Rockwell
Rockwell 5,6
Rockwell (EN ISO 18265 only)
Rockwell
Knoop (ASTM E140 only)
Shore (ASTM E140 only)
Tensile strength
|
HV
HB
HRB
HRC
HRE
HRF
HRA
HRD
HR45N
HK
HS
MPa
|
10 – 1999 (9999)
76 – 618
41 – 105
20,3 – 68
70 – 108,5
82,6 – 115,1
60,7 – 8
40,3 – 76,9
19,9 – 75,4
87 – 920
34,2 – 97,3
255 – 2180
|
|
Measurement
uncertainty*
|
< 4 % (HV5, HV 10). For other test
loads and ranges see table below.
|
|
Relative
repeatability*
|
< 5 % (HV5, HV 10). For other test
loads and ranges see table below.
|
* exceeds DIN 50159, dependent on test load
and range (see table below). Specifications are valid for 5 measurements
usingVickers reference blocks and according to test conditions given in
standard DIN 50159.
NewSonic SonoDur3超聲波硬度計(jì)
Scope of delivery
SonoDur3, Hardness Tester with Data Logger
and Data Export, Data Transfer to PC (USB, WLANor Bluetooth resp.), incl.
SONO3-NG, Power Source, SONO2-NG/USB USB-Cable, SONO2-HM, ca. 1,5 mProbe
Connection Cable, SONO3-TK-1, Transportation Case, SONO-CD, Product-USB-Stick,
SONO3Protect, Protection Foils
Mandatory
accessory
Motor probe or handheld probe respectively
Other accessories
Reference blocks (MPA, company
certificate),test stands, guiding harnesses, auxiliary SW.
Example for
configurations
溫馨提示:深圳泰立儀器代理銷售德國(guó)NewSonic超聲波硬度計(jì),因美國(guó)GE 硬度計(jì)MIC10已停產(chǎn),
好消息:德國(guó)NewSoni新一代超聲波硬度計(jì)SonoDur3設(shè)備完全替代美國(guó)GE硬度計(jì)并兼容已停產(chǎn)老款探頭.
提供可連接 MIC10 和 MIC10DL 的硬度計(jì)
SONO-201 負(fù)載力 10N(1kg)
SONO-205 負(fù) 載力 49N(5kg)
SONO-2010 負(fù)載力 98N(10kg)
注意選用此系列探頭搭配MIC10及MIC10DL需要更換以下探頭線SONO-N1050才可以正常使用:
MIC10探頭線
SONO-M1 (0.5米 GE電動(dòng)探頭配合支架適用)
SONO-M10M (1.5米 GE電動(dòng)探頭)
SONO-M1C1050 1.5米 ( GE手動(dòng)探頭適用)